tct_setup
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| tct_setup [2024/12/11 20:05] – [Procedure] prezae | tct_setup [2024/12/11 21:40] (current) – [Software] prezae | ||
|---|---|---|---|
| Line 79: | Line 79: | ||
| - Activate the laser using the software on the computer and verify using the oscilloscope that a signal can be observed in at least one of the pads when triggering to the laser' | - Activate the laser using the software on the computer and verify using the oscilloscope that a signal can be observed in at least one of the pads when triggering to the laser' | ||
| - Manually locate the four corners by adjusting the position of the sensor using the software. Increasing the position in the positive x-direction shifts the plate towards the right side of the box, and increasing the position in the positive y-direction shifts the plate towards the back of the box. Once at the corner of an active region, the observed waveform amplitude will decrease and totally disappear. | - Manually locate the four corners by adjusting the position of the sensor using the software. Increasing the position in the positive x-direction shifts the plate towards the right side of the box, and increasing the position in the positive y-direction shifts the plate towards the back of the box. Once at the corner of an active region, the observed waveform amplitude will decrease and totally disappear. | ||
| - | - {{ prosedure_1.png? | ||
| - Once the four corners have been located, shine the laser somewhere in the middle of one of the four quadrants. There is a macro script to run which adjusts the z position (height) of the laser while measuring the voltage. Using this, we can find out the optimum distance the laser should be from the sensor. | - Once the four corners have been located, shine the laser somewhere in the middle of one of the four quadrants. There is a macro script to run which adjusts the z position (height) of the laser while measuring the voltage. Using this, we can find out the optimum distance the laser should be from the sensor. | ||
| + | - Alternatively, | ||
| + | - {{ prosedure_1.png? | ||
| ==== Trigger ==== | ==== Trigger ==== | ||
| Line 97: | Line 98: | ||
| To run the script, enter the following command into the terminal: | To run the script, enter the following command into the terminal: | ||
| python3 TheSetup.py | python3 TheSetup.py | ||
| - | Next, to access the graphical interface of the motors, locate the script " | + | To access the graphical interface of the motors, locate the script " |
| code/ | code/ | ||
| Then, enter the following command into the terminal to run the script: | Then, enter the following command into the terminal to run the script: | ||
tct_setup.1733943922.txt.gz · Last modified: by prezae