This is an old revision of the document!
We have two power supply units, both controlled by a LabVIEW script via the server. We use a Low Voltage Power Supply Unit (LVPSU), the HMP 2030 by Rohde & Schwarz. The LVPSU powers the LTA with 12 V and 1-2 A and supplies the base power for the CCD. For a High Voltage Power Supply Unit (HVPSU), we use the Keithley 2470 SourceMeter. It is used to deplete the CCD of any free charge, using around 50-70 V. It has very high precision, capable of measuring currents down to femtoamperes. This precision is needed to test for defects in CCDs.
Current and voltage range of the HMP2030
| Voltage Range | Current (A) | Power Management | Max Current at 32 V (A) |
|---|---|---|---|
| 0 - 16 | 5 | Standard | - |
| 16 - 32 | 2.5 | Intelligent Power Management | 2.5 |
The VME crate provied the following outputs:
| Output | Power (V) | max Current (A) |
|---|---|---|
| 1 | +3.3 | 25 |
| 2 | +5 | 25 |
| 3 | -12 | 10 |
| 4 | +12 | 10 |
| User Input -V1 | -15 | Set by HMP2030 |
| User Input +V1 | +15 | |
| User Input -V2 | -30 | |
| User Input +V2 | N/C | |
To generate this voltages/currents it uses a Elma VME backplane (Article Number: 32220-204-0B60) 4 slot 6 HE VME64x 4 Slots with P0, with M3/M4 power bolts board (Link). I has the following features: