 ===== Depressurization Checklist ===== ===== Depressurization Checklist =====
  
-1)      Ensure that cryocooler is off! +  - Ensure that cryocooler is off and cryostat is at room temps
-2) Turn on main power switch on HiCube Eco 80 +  Turn on main power switch on HiCube Eco 80 
-3) Wait for DCU to complete self-test of system +  Wait for DCU to complete self-test of system 
-4) Check if venting is off under parameter 012 “Enable vent” => off +  Check if venting is off under parameter 012 “Enable vent” => off 
-5) Turn on roughing pump: Select parameter 010 “Pumping station” => on  +  Turn on roughing pump: Select parameter 010 “Pumping station” => on  
-6) Once we reach pressures !!< 5 mbar!!, select parameter 023 “ motor pump” and turn on turbomolecular pump +  Once we reach pressures !!< 5 mbar!!, select parameter 023 “ motor pump” and turn on turbomolecular pump 
-7)      Ideally, we want pressure to drop to lower 10^(-6) mbar range +  Ideally, we want pressure to drop to lower 10^(-6) mbar range  
 + 
 + 
 + 
 + 
 +===== Pressurization Checklist ===== 
 + 
 +  - Turn cryocooler off and wait for temperatures to reach ~ 280 K 
 +  - Turn off Turbomolecular pump and wait for RPMs to go to zero 
 +  - Turn off roughing pump 
 +  - Vent the system 
  
 ===== TPG 361 ===== ===== TPG 361 =====