User Tools

Site Tools


psi:psi

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
psi:psi [2017/01/11 18:44] – [Testing procedure] leacpsi:psi [2017/01/17 11:48] (current) – [List of Calibrations] leac
Line 11: Line 11:
       * Optical fibers (tkFEC, pxFEC, FED 1-5)       * Optical fibers (tkFEC, pxFEC, FED 1-5)
   - Verify that cooling sytem is running   - Verify that cooling sytem is running
-      * Linke to Malte's twiki: +      * See Malte's twiki: https://twiki.cern.ch/twiki/bin/view/CMS/CO2LEANOperationGuide
   - Verify that temperature and humidity reading are as expected   - Verify that temperature and humidity reading are as expected
   - Power on CCU and opto power   - Power on CCU and opto power
Line 41: Line 41:
   prints list of all i2c devices    prints list of all i2c devices 
  
-If you do not see the expected output, try to powercycle the CCU and opto power.  +  * If you do not see the expected output, try to powercycle the CCU and opto power.  
-If scanringedevice crashes, try to disconnect and reconnect the fibers going to the TKFEC. +  If scanringedevice crashes, try to disconnect and reconnect the fibers going to the TKFEC. 
  
  
Line 64: Line 64:
      
      
 +=== How to analyze PixelAlive === 
 +  
 +  $BUILD_HOME/pixel/PixelAnalysisTools/test/bin/linux/x86_64_slc6/PixelAnalysis.exe $BUILD_HOME/pixel/PixelAnalysisTools/test/configuration/PixelAliveAnalysis.xml yourrunnur
 +  
 +Edit xml file to include the FEDs of interest.  
 +===== List of Calibrations =====   
 +
 +  TBMDelay25 --> updatePortcard.
 +  if needed: TBMPLLDelay --> updateTbm for individual modules if needed.
 +  VcthrCalDel with HV for L1 modules (need to change detconfig) --> updateRoc. 
 +  PixelAlive with HV (for full detconfig).
 +  PixelAlive without HV. 
 +  
 +===== How to go to the next sector ===== 
 +  - Connect module power and fibers
 +  - Change POS configuration files to appropriate sector
 +
 +  cd TriDAS_r45/pixel/config
 +  cd detconfig/1/
 +  cp detectconfig.dat_sector detconfig.dat
 +  cd ../../nametranslation/1/
 +  cp translation.dat_sector translation.dat
 +  cd ../../portcardmap/1/
 +  cp portcardmap.dat_sector portcardmap.dat
 +  
 +  
 +===== How to telnet to CAEN ===== 
 +
 +  telnet 129.129.202.207 1527
 +  
 +  Login using user: user, password: user
 ===== Elog =====   ===== Elog =====  
  
Line 70: Line 101:
   https://elog.physik.uzh.ch:8080/PixelPOS/   https://elog.physik.uzh.ch:8080/PixelPOS/
      
-  Enter sector, run number and calibration and additional information if needed.+  *   For each calibration: enter sector, run number and calibration type and additional information if needed. 
 +  *   For each sector, enter a summary of working/non-working components.  
 + 
 +===== How to start up temperature sensor reading =====   
 + 
 +Login to the PC: user: PC7530\cms, password: Pixel2017 
 +   
 +  Start LabView2015 
 +  Select project TempReadout 
 +  Open file TempReadout_pc.vi 
 +  Click play in the upper left corner 
 +   
 +  Before exiting the program, click stop in the upper right corner!
 ===== mTCA specifics ===== ===== mTCA specifics =====
   * To load firmware after power cycle of the crate see: https://wiki.physik.uzh.ch/cms/phase1:phase1:mtcafull   * To load firmware after power cycle of the crate see: https://wiki.physik.uzh.ch/cms/phase1:phase1:mtcafull
psi/psi.1484156648.txt.gz · Last modified: 2017/01/11 18:44 by leac