| ===== Depressurization Checklist ===== | ===== Depressurization Checklist ===== | ||
| - | 1) | + | - Ensure that cryocooler is off and cryostat is at room temps! |
| - | 2) Turn on main power switch on HiCube Eco 80 | + | |
| - | 3) Wait for DCU to complete self-test of system | + | |
| - | 4) Check if venting is off under parameter 012 “Enable vent” => off | + | |
| - | 5) Turn on roughing pump: Select parameter 010 “Pumping station” => on | + | |
| - | 6) Once we reach pressures !!< 5 mbar!!, select parameter 023 “ motor pump” and turn on turbomolecular pump | + | |
| - | 7) | + | |
| + | |||
| ===== Pressurization Checklist ===== | ===== Pressurization Checklist ===== | ||
| - | 1) | + | - Turn cryocooler off and wait for temperatures to reach ~ 280 K |
| - | 2) Turn off Turbomolecular pump and wait for RPMs to go to zero | + | |
| - | 3) Turn off roughing pump | + | |
| - | 4) Vent the system | + | |
| + | |||
| ===== TPG 361 ===== | ===== TPG 361 ===== | ||
| The pressure within the Cryostat is monitored with a **Single Gauge TPG361** from **Pfeiffer Vaccum**, which uses a PKR 251 FPM-sealed, DN 25 ISO-KF gauge sensor. The maximum pressure applies to inert gases and temperatures below 55 °C and is corrosion resistant. The TPG361 is designed for installation to a 19“ rack chassis adapter according to DIN 41494 and provides serial communication via R485 (24V, 5-pin circular), USB and Ethernet interfaces.It is used to measure the pressure inside the chamber. | The pressure within the Cryostat is monitored with a **Single Gauge TPG361** from **Pfeiffer Vaccum**, which uses a PKR 251 FPM-sealed, DN 25 ISO-KF gauge sensor. The maximum pressure applies to inert gases and temperatures below 55 °C and is corrosion resistant. The TPG361 is designed for installation to a 19“ rack chassis adapter according to DIN 41494 and provides serial communication via R485 (24V, 5-pin circular), USB and Ethernet interfaces.It is used to measure the pressure inside the chamber. | ||