| As a solenoid valve funtions the **0211** from Bürkert, | As a solenoid valve funtions the **0211** from Bürkert, | ||
| - | ===== Starting Procedure | + | ===== Depressurization Checklist |
| + | |||
| + | - Ensure that cryocooler is off and cryostat is at room temps! | ||
| + | - Turn on main power switch on HiCube Eco 80 | ||
| + | - Wait for DCU to complete self-test of system | ||
| + | - Check if venting is off under parameter 012 “Enable vent” => off | ||
| + | - Turn on roughing pump: Select parameter 010 “Pumping station” => on | ||
| + | - Once we reach pressures !!< 5 mbar!!, select parameter 023 “ motor pump” and turn on turbomolecular pump | ||
| + | - Ideally, we want pressure to drop to lower 10^(-6) mbar range | ||
| + | |||
| + | |||
| + | |||
| + | |||
| + | ===== Pressurization Checklist ===== | ||
| + | |||
| + | - Turn cryocooler off and wait for temperatures to reach ~ 280 K | ||
| + | - Turn off Turbomolecular pump and wait for RPMs to go to zero | ||
| + | - Turn off roughing pump | ||
| + | - Vent the system | ||
| - | First turn the vacuum pump on, and then make sure that the Turbomolacularpump is turned off. Only start the vacuum pump when you are sure that the Turbo is off, then start the compressor pump, when the pressure reaches the lower single digits (5 mbar >) turn the Turbomolacularpump on. | ||
| ===== TPG 361 ===== | ===== TPG 361 ===== | ||
| The pressure within the Cryostat is monitored with a **Single Gauge TPG361** from **Pfeiffer Vaccum**, which uses a PKR 251 FPM-sealed, DN 25 ISO-KF gauge sensor. The maximum pressure applies to inert gases and temperatures below 55 °C and is corrosion resistant. The TPG361 is designed for installation to a 19“ rack chassis adapter according to DIN 41494 and provides serial communication via R485 (24V, 5-pin circular), USB and Ethernet interfaces.It is used to measure the pressure inside the chamber. | The pressure within the Cryostat is monitored with a **Single Gauge TPG361** from **Pfeiffer Vaccum**, which uses a PKR 251 FPM-sealed, DN 25 ISO-KF gauge sensor. The maximum pressure applies to inert gases and temperatures below 55 °C and is corrosion resistant. The TPG361 is designed for installation to a 19“ rack chassis adapter according to DIN 41494 and provides serial communication via R485 (24V, 5-pin circular), USB and Ethernet interfaces.It is used to measure the pressure inside the chamber. | ||
| * Front panel Operation: {{ : | * Front panel Operation: {{ : | ||
| * Supported Commands:{{ : | * Supported Commands:{{ : | ||
| + | * Data Format: 1 start bit, 8 data bits, no parity bit, 1 stop bit, no hardware handshake | ||
| ^ Characteristic | ^ Characteristic | ||
| Line 39: | Line 58: | ||
| ===== Vacuum system configuration ===== | ===== Vacuum system configuration ===== | ||
| The Pfeiffer Vacuum' | The Pfeiffer Vacuum' | ||
| - | ^ Subsystem | + | ^ Subsystem |
| - | | ::: | + | | | Name | Model Nr. | Manuals |
| - | | Roughening Pump | MVP 015-2 | Not visible | + | | Roughening Pump | MVP 015-2 | Not visible |
| - | | Turbomulecular | + | | Turbomolecular |
| - | | Power Supply Unit | TPS110 | + | | Power Supply Unit | TPS110 |
| - | | Control Unit | TC110 | PM C01 790 A | + | | Control Unit | TC110 | PM C01 790 A | {{ : |
| - | | Venting Valve | | | Power Failure Vent | PM Z01 330 A | | + | | Venting Valve | Power Failure Vent |
| - | | Cooling Fan | Air cooling Kit 24V | PM Z01 300 A | + | | Cooling Fan | Air cooling Kit 24V | PM Z01 300 A | {{ : |
| - | | Front control Panel | DCU002 | + | | Front control Panel | DCU002 |
| + | | Gauge | TPG261 | ||
| + | | Vacuum Sensor | ||